The Microscope – Volume 73, Third Quarter 2026
IN THIS ISSUE
On the cover
Physical arrangement of the DCMI components and accessories and setup using a Leitz Ortholux I polarized light microscope. See Differential Central Mask Illumination (DCMI): A Non-Interferometric, Tunable Phase-Gradient Contrast Method for Transmitted-Light Microscopy, page 99.(Courtesy of William A. Mark)
Editorial | H
Gary J. LaughlinThe Microscope 73:3, p. ii, 2026https://doi.org/10.59082/BXDR8013
Excerpt:
Differential Central Mask Illumination (DCMI): A Non-Interferometric, Tunable Phase-Gradient Contrast Method for Transmitted-Light Microscopy
William A. MarkThe Microscope 73:3, pp. 99–111, 2026https://doi.org/10.59082/XCBU1533
Abstract: Differential Central Mask Illumination (DCMI) is a transmitted-light contrast method intended to make phase-gradient information easier to see in transparent and weakly absorbing specimens. The method uses a simple illumination-side optical modulation filter stack: a polarizer, an oriented birefringent step wedge, a central circular oblique lighting (COL) mask made from polarizing film, and an analyzer above the objective. In combination, these elements produce an asymmetric weighting of light in the objective back focal plane. The resulting image has a directional, relief-like appearance related in some respects to differential interference contrast (DIC), however it is not produced by prism-based beam shearing or recombination.
The method was tested on diatoms, crystals, stained plant tissue, hair, starch granules, and wet-mount crystal preparations. Back-focal-plane observations show the development of the directional progressive imbalance of asymmetry as the polarizers, wedge, and central mask are introduced. Image comparisons with Zeiss DIC (Gen 1) show that DCMI does not reproduce DIC exactly; however, it can give useful structural contrast in many of the same types of transparent specimens. Fast Fourier transform (FFT) analysis was used as a descriptive check on the recorded images. The main FFT result is not improved resolution, but redistribution of image energy and increased directional anisotropy in selected spatial-frequency bands. DCMI is therefore best presented as a low-cost, adaptable, and tunable contrast method that combines features of COL, oblique illumination, and polarization microscopy while remaining optically distinct from DIC.
Shelf Life for the Gold Chloride in Phosphoric Acid Microcrystal Test Reagent Used for Testing Methamphetamine by Polarized Light Microscopy
David M. Northrop
The Microscope 73:3, pp. 112–114, 2026https://doi.org/10.59082/FMER6728
Abstract: Microchemical and microcrystal tests are valuable tools for forensic chemists conducting seized drug analysis. Several microcrystal tests using the polarized light microscope (PLM) are available that provide simple, quick, and reliable information about the contents of a suspected seized drug exhibit. Gold chloride in phosphoric acid (gold chloride/phosphoric acid reagent) is a reagent for microcrystal tests that can be used to indicate the presence of methamphetamine in an exhibit.
Quality control measures to verify that the reagent performs properly are typically conducted at regular intervals. In this study, a gold chloride/phosphoric acid reagent used for microcrystal tests was verified using a known, primary methamphetamine standard, once a month, for 25 years. Appropriate microcrystal test results were obtained throughout the study, with the result being that the shelf life, i.e., length of time that the gold chloride/phosphoric acid reagent, can be stored at room temperature before it becomes too old to be used, is at least 25 years, or more. This information can assist in establishing verification intervals.
Water-Driven Panning as an Alternative to Heavy Liquid Separation
John Palenik
The Microscope 73:3, pp. 115–120, 2026
https://doi.org/10.59082/GTNP9453
Abstract: Most of the world’s sands are made up of “light” silicates, such as quartz and feldspars. Because of their frequency and abundance, these minerals are generally less characteristic of a sand’s identity than heavier minerals. Separating sand samples into heavy and light mineral fractions then becomes an important preparatory step towards sand characterization, such as geo-sourcing analysis.
The standard process of heavy liquid separation (HLS) is effective but limited to a laboratory environment, since the chemicals used in this process (e.g. bromoform) are difficult to obtain, expensive, and bad for the environment. In this project, the technique of water-driven panning is explored as an alternative method for isolating heavy minerals.
Samples of beach sand were collected, processed, and analyzed to determine the effectiveness of this panning technique. These included control samples, fully and partially panned sands (including magnetically-separated samples), and heavy mineral deposits. Particle counts of 1,000 grains of sand from each of the samples were conducted, logging the numbers of light minerals, heavy-transparent minerals, and heavy-opaque minerals in each sample. Individual minerals were characterized through a combination of polarized light microscopy and microchemical tests. The results showed that compared to the control samples, panning dramatically reduced the abundance of light minerals (from 97% to 5%) and greatly concentrated heavy minerals. It is thus demonstrated that water-driven panning could be a cheap, effective, and environmentally friendly field tool for preparing sands for analysis.
The Suitability of Polar Bear Hair as an Inspiration for Biomimicry
K. Paul Kirkbride and Silvana R. Tridico
The Microscope 73:3, pp. 121–126, 2026https://doi.org/10.59082/APWY2734
Abstract: A substantial number of articles in the technical literature and the Internet state that polar bear hairs are “hollow.” It is further stated or implied that the “hollowness” of these hairs is unique or at least unusual, that it confers special thermal regulation capabilities to the animal, and that these hairs can act as a model for the development of thermally-insulative biomimetic materials. However, this strand of biomimetic materials research could be misdirected, because the purported unique property of polar bear guard hairs – hollowness - is not correct and seems to arise, in part, from misinterpretation of transmitted light microscopical images of the medullae of those hairs.
Furthermore, other properties of polar bear hairs purported to arise from cold adaptation, such as scales and radial pores, are also not cold adaptations.
It appears to be the case that other properties of the polar bear pelage, such as the high ratio of under hairs to guard hairs, is much more important regarding thermal insulation.
An Obvious Illuminator for Dispersion Staining
Oppenheimer GoldbergThe Microscope 73:3, pp. 127–130
Abstract: The ideal illuminator for dispersion staining would be a continuous white light laser. The features making the ideal are parallelism of the beam, small beam diameter, and high intensity. The illuminator chosen as a substitute for the unavaialable ideal most come as close to these features as possible. The obvious substitute is based on fiber optics to transfer the light from an unfrosted, small filament lamp, to a point directly below the substage aperture diaphram.
Critical Focus | Mad Cows, Kuru, and a New World Record
Brian J. Ford
The Microscope 73:3, pp. 131–143, 2026https://doi.org/10.59082/SRDJ8184
Excerpt: Thirty years ago, humanity seemed to face an existential crisis. An insidious new disease had dramatically emerged. It caused a progressive breakdown in the brain; it was invariably fatal, there was no treatment, no vaccine, and no cure. The pathogen was so primitive that it lacked any trace of genetic material; yet it was so advanced that it caused an infection that could threaten the survival of humanity. The media called it Mad Cow Disease; we knew it as Bovine Spongiform Encephalopathy (BSE), a brain infection caused by the simplest kind of pathogen imaginable, smaller than viruses – a prion.
Little was known about it, and the public was completely in the dark. It fell to me to summarize what we knew in a book, and that was to set a world record in book publishing. Starting from cold, it was a nonfiction work that was researched, organized, written, illustrated, indexed, edited, printed, and bound within a month. Nothing like that in the world of publishing had ever been done before, and nothing like it has been done since.
Afterimage | Under pressure
Sebastian Sparenga — McCrone Research InstituteThe Microscope 73:3, p. 144, 2026
Malonic acid thin-film fusion prep "lens crystal" made by applying pressure to the center of the coverslip with forceps as the slide cools; original magnification is 40×, circularly polarized light.
Malonic acid thin-film fusion prep "lens crystal" made by applying pressure to the center of the coverslip with forceps as the slide cools; original magnification is 40×, circularly polarized light.
Copyright © 202 Microscope Publications, Division of McCrone Research Institute. All rights reserved.